Package: metrawatt_metrahit_1x_1yr_ver_9100.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

METRAWATT METRAHit 12S: (1 year) CAL VER /9100

METRAWATT METRAHit 13S: (1 year) CAL VER /9100

METRAWATT METRAHit 14A: (1 year) CAL VER /9100

METRAWATT METRAHit 14S: (1 year) CAL VER /9100

METRAWATT METRAHit 15S: (1 year) CAL VER /9100

Date: 2003-07-14

Revision: 1.4

Automated Reference: Datron 9100

Model(s) Covered: 12S, 13S, 14A, 14S, 15S

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

CAPACITANCE

29 nF; TUR 2.2566

290 nF; TUR 3.7603

2.9 uF; TUR 3.7603

29 uF; TUR 2.7350

Source: GOSSEN-METRAWATT

Specifications in Operating Instructions, 11/2.97

METRA Hit 12S ... 16S

Analog Digital Multimeter

GOSSEN-METRAWATT Quality Control

Test Certificate for METRAHit 15S

Serial No. M49963645

Compatibility: MET/CAL 7.01 or later

Additional Equipment Required: None

The average run time during testing was 15 minutes.

METRAWATT METRAHit 12S: CAL ADJ /9100

METRAWATT METRAHit 13S: CAL ADJ /9100

METRAWATT METRAHit 14A: CAL ADJ /9100

METRAWATT METRAHit 14S: CAL ADJ /9100

METRAWATT METRAHit 15S: CAL ADJ /9100

Date: 2003-07-14

Revision: 1.1

Automated Reference: Datron 9100

Model(s) Covered: 12S, 13S, 14A, 14S, 15S

Source: GOSSEN-METRAWATT

Specifications in Operating Instructions, 11/2.97

METRA Hit 12S ... 16S

Analog Digital Multimeter

GOSSEN-METRAWATT Quality Control

Test Certificate for METRAHit 15S

Serial No. M49963645

Compatibility: MET/CAL 7.01 or later

Additional Equipment Required: None

METRAWATT METRAHit 16S: (1 year) CAL VER /9100

Date: 2003-07-14

Revision: 1.4

Automated Reference: Datron 9100

Model(s) Covered: 16S

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

CAPACITANCE

29 nF; TUR 2.2566

290 nF; TUR 3.7603

2.9 uF; TUR 3.7603

29 uF; TUR 2.7350

Source: GOSSEN-METRAWATT

Specifications in Operating Instructions, 11/2.97

METRA Hit 12S ... 16S

Analog Digital Multimeter

GOSSEN-METRAWATT Quality Control

Test Certificate for METRAHit 15S

Serial No. M49963645

Compatibility: MET/CAL 7.01 or later

Additional Equipment Required: None

The average run time during testing was 15 minutes.

METRAWATT METRAHit 18S: (1 year) CAL VER /9100

Date: 2003-07-14

Revision: 1.2

Automated Reference: Datron 9100

Model(s) Covered: 18S

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

AC CURRENT

280 uA @ 50 Hz; TUR 3.4274

CAPACITANCE

2.7 nF; TUR 1.5152

28.5 nF; TUR 3.2035

285 nF; TUR 3.1527

2.85 uF; TUR 2.4615

8730 uF; TUR 3.0075

Source: Portocal II Procedure Version 1.70

Compatibility: MET/CAL 7.01 or later

Additional Equipment Required: None

The average run time during testing was 15 minutes.