Package: metrawatt_metrahit_2x_1yr_ver_5520.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

METRAWATT METRAHit 22S/M: (1 year) CAL VER /5520

METRAWATT METRAHit 23S: (1 year) CAL VER /5520

METRAWATT METRAHit 24S: (1 year) CAL VER /5520

METRAWATT METRAHit 25S: (1 year) CAL VER /5520

METRAWATT METRAHit 26S/M: (1 year) CAL VER /5520

Date: 2003-07-15

Revision: 1.6

Automated Reference: Fluke 5520A

Model(s) Covered: 22S/M, 23S, 24S, 25S, 26S/M

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

CAPACITANCE

2.85 nF; TUR 1.6563

28.5 nF; TUR 2.2381

28.5 uF; TUR 2.9870

Source: GOSSEN-METRAWATT

Specifications in Operating Instructions, 2/8.99

Analog Digital Multimeter with Signal Generator

22S/M, 23S, 24S, 25S, 26S/M

DKD-K-19701 Calibration Certificate for METRAHit 23S

Serial number JK3406, 15.11.1999

Compatibility: MET/CAL or 5500/CAL 7.01 or later

Additional Equipment Required: None

The average run time during testing was 20 minutes.

METRAWATT METRAHit 28S: (1 year) CAL VER /5520

METRAWATT METRAHit 29S: (1 year) CAL VER /5520

Date: 2003-07-15

Revision: 1.6

Automated Reference: Fluke 5520A

Model(s) Covered: 28S, 29S

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

DC CURRENT

280 uA; TUR 3.8246

CAPACITANCE

2.85 nF; TUR 1.6563

28.5 nF; TUR 2.2381

28.5 uF; TUR 2.9870

Source: GOSSEN-METRAWATT

Specifications from website, 6/03

DKD-K-19701 Calibration Certificate for METRAHit 23S

Serial number JK3406, 15.11.1999

Compatibility: MET/CAL or 5500/CAL 7.01 or later

Additional Equipment Required: None

The average run time during testing was 20 minutes.