Package: quadtech_sentry_10_plus_1yr_ver_5320_5320load_dmm.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

QuadTech Sentry 10 Plus: (1 yr) VER /5320,5320LOAD,DMM

Date: 2007-07-17 13:55:29

Revision: 1.3

Automated Reference: Digital Multimeter

Automated Reference: Fluke 5320A

Manual Reference: Fluke 5320A-LOAD

Model(s) Covered:

Sentry 10 Plus

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

AC Voltage

- 1.000 kV TUR 1.5000

- 2.000 kV TUR 1.6667

- 5.000 kV TUR 1.8333

Source:

QuadTech Sentry Plus Series

Hipot Testers

Instruction Manual

Form 150697/A3, February 2004

Sentry Plus B & A Data Sheet

Form 150701/A1

Compatibility:

MET/CAL 7.20d or later

Additional Equipment Required:

- Digital Multimeter, suitable for use with configured

DMM, 0 to 40 mA, 0. % accuracy

- 5 kV test leads: 3 required

single banana (m) to single banana (m)

- Adapter: Clip lead to single banana (m)

The average run time during testing was 15 minutes.