Package: radian_research_rm-10_1yr_ver_6105a_e_80a.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Radian Research RM-10: (1 year) CAL VER /6105A+E+80A

Date: 2010-05-12 12:45:25

Revision: 1.1

Automated Reference: Fluke 6105A

Model(s) Covered:

RM-10

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1 based on 6105A 1 year specifications.

120, 240, 480, & 600 V at:

25 A, 0° TUR 2.25

25 A, 60° Lag TUR 1.33

30 A, 0° TUR 2.45

30 A, 60° Lag TUR 1.37

40 A, 0° TUR 2.75

40 A, 60° Lag TUR 2.46

45 A, 0° TUR 2.87

45 A, 60° Lag TUR 2.55

50 A, 0° TUR 2.97

50 A, 60° Lag TUR 2.62

Compatibility:

MET/CAL or 7.3.9 or later

Additional Equipment Required:

None

The average run time during testing was 54 minutes.