Package: radian_research_rm-10_1yr_ver_6105a_e_80a.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Radian Research RM-10: (1 year) CAL VER /6105A+E+80A, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Radian Research RM-10: (1 year) CAL VER /6105A+E+80A
Date: 2010-05-12 12:45:25
Revision: 1.1
Automated Reference: Fluke 6105A
Model(s) Covered:
RM-10
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1 based on 6105A 1 year specifications.
120, 240, 480, & 600 V at:
25 A, 0° TUR 2.25
25 A, 60° Lag TUR 1.33
30 A, 0° TUR 2.45
30 A, 60° Lag TUR 1.37
40 A, 0° TUR 2.75
40 A, 60° Lag TUR 2.46
45 A, 0° TUR 2.87
45 A, 60° Lag TUR 2.55
50 A, 0° TUR 2.97
50 A, 60° Lag TUR 2.62
Compatibility:
MET/CAL or 7.3.9 or later
Additional Equipment Required:
None
The average run time during testing was 54 minutes.