Package: rs_fsh6_1yr_ver_rs232_9640_e441x_swpg.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Rohde & Schwarz FSH26:(1 yr) VER RS-232 /9640,4418,SWPG, Revision: 1.1
Rohde & Schwarz FSH6: (1 yr) VER RS-232 /9640,4418,SWPG, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Rohde & Schwarz FSH26:(1 yr) VER RS-232 /9640,4418,SWPG
Rohde & Schwarz FSH6: (1 yr) VER RS-232 /9640,4418,SWPG
Date: 2007-04-16 15:36:56
Revision: 1.1
Automated Reference: Agilent E4418B
Automated Reference: Agilent E9304A
Automated Reference: Fluke 9640A
Automated Reference: Fluke 9640A-50
Automated Reference: Sweep Generator
Manual Reference: Agilent 11667A
Model(s) Covered:
FSH6, FSH26
Source(s):
Rohde & Schwarz Handheld Spectrum Analyzer R&S FSH Service Manual,
1145.5973.82-12 (December 2006)
USAF T.O. 33K3-4-3494-1, Technical Manual, Calibration Procedure
For Spectrum Analyzer, FSH6, dated 28 February 2006
Compatibility:
MET/CAL 7.20c or later
Flexible Standards Required:
HF Signal Generator (HFSG):
Frequency Range: 500 MHz
Accuracy: ±1 MHz
Amplitude Range: 0 dBm
Accuracy: NA
Phase Noise:
<-100 dBc/Hz @ 10 kHz offset
<-110 dBc/Hz @ 100 kHz offset
<-130 dBc/Hz @ 1 MHz offset
Synthesized Sweeper (SWPG):
Frequency Range: 10 MHz to 6 GHz
Accuracy: ±0.25%
Amplitude Range: -35 dBm to +6 dB
Accuracy: NA
SSB Phase Noise:
<-70 dBc/Hz @ 50 kHz offset
Additional Equipment Required:
Power Splitter, DC-6 GHz, HP/Agilent 11667A or equivalent
Termination, BNC (m), 50 Ohms, DC-6 GHz, HP/Agilent 909C or
equivalent
Cable, coaxial, 36", 50 Ohms, BNC (m)-(m), RG-58/U
Rohde & Schwarz FSH-Z33 AC Power Supply for the FSH
Rohde & Schwarz FSH-Z34 Optically Isolated Interface Cable
Fluke Adapter Kit, Part No. 689486 or equivalent. Includes DB9(f)
to DB9(f) null modem adapter and DB9(m) to DB9(m) gender changer
The average run times during testing were:
20 minutes (TAR mode)