Package: rs_fsh6_1yr_ver_rs232_9640_e441x_swpg.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Rohde & Schwarz FSH26:(1 yr) VER RS-232 /9640,4418,SWPG

Rohde & Schwarz FSH6: (1 yr) VER RS-232 /9640,4418,SWPG

Date: 2007-04-16 15:36:56

Revision: 1.1

Automated Reference: Agilent E4418B

Automated Reference: Agilent E9304A

Automated Reference: Fluke 9640A

Automated Reference: Fluke 9640A-50

Automated Reference: Sweep Generator

Manual Reference: Agilent 11667A

Model(s) Covered:

FSH6, FSH26

Source(s):

Rohde & Schwarz Handheld Spectrum Analyzer R&S FSH Service Manual,

1145.5973.82-12 (December 2006)

USAF T.O. 33K3-4-3494-1, Technical Manual, Calibration Procedure

For Spectrum Analyzer, FSH6, dated 28 February 2006

Compatibility:

MET/CAL 7.20c or later

Flexible Standards Required:

HF Signal Generator (HFSG):

Frequency Range: 500 MHz

Accuracy: ±1 MHz

Amplitude Range: 0 dBm

Accuracy: NA

Phase Noise:

<-100 dBc/Hz @ 10 kHz offset

<-110 dBc/Hz @ 100 kHz offset

<-130 dBc/Hz @ 1 MHz offset

Synthesized Sweeper (SWPG):

Frequency Range: 10 MHz to 6 GHz

Accuracy: ±0.25%

Amplitude Range: -35 dBm to +6 dB

Accuracy: NA

SSB Phase Noise:

<-70 dBc/Hz @ 50 kHz offset

Additional Equipment Required:

Power Splitter, DC-6 GHz, HP/Agilent 11667A or equivalent

Termination, BNC (m), 50 Ohms, DC-6 GHz, HP/Agilent 909C or

equivalent

Cable, coaxial, 36", 50 Ohms, BNC (m)-(m), RG-58/U

Rohde & Schwarz FSH-Z33 AC Power Supply for the FSH

Rohde & Schwarz FSH-Z34 Optically Isolated Interface Cable

Fluke Adapter Kit, Part No. 689486 or equivalent. Includes DB9(f)

to DB9(f) null modem adapter and DB9(m) to DB9(m) gender changer

The average run times during testing were:

20 minutes (TAR mode)