R&S FSP7 (FSP-B9): (1 yr) VER /LPNX,4419,LFCTR,LO,SWPG

Date: 2011-07-01

Revision: 1.1

Automated Reference: Agilent E4418B

Automated Reference: Agilent E4419B

Automated Reference: Agilent E9304A

Automated Reference: Fluke 9640A

Automated Reference: Fluke 9640A-50

Automated Reference: Fluke 9640A-LPNX

Automated Reference: HP 8481A

Automated Reference: HP 8482A

Automated Reference: LF Counter

Automated Reference: Local Oscillator

Automated Reference: Low Frequency Counter

Automated Reference: Sweep Generator

Manual Reference: Agilent 11667A

Model(s) Covered:

FSP7

Source(s):

R&S® Service Manual Instrument, Spectrum Analyzer, R&S® FSP3,

R&S® FSP7, R&S® FSP13, R&S® FSP30, R&S® FSP31, R&S® FSP40, Part No.

1164.4556.82-01-

R&S® FSP Spectrum Analyzer Data Sheet, Version 06.00, May 2008

Compatibility:

MET/CAL 7.3 SP1 (7.3.36) or later

Flexible Standard(s) Required:

Synthesized Sweep Generator (SWPG):

Frequency Range: 10 MHz to 10 GHz, CW

Accuracy: ±0.02% (Internal); External Reference capable

Amplitude Range: -40 dBm to +16 dBm

Resolution: 0.02 dB

SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset

Example: HP 83630B or equivalent

Local Oscillator (LO):

Frequency Range: 10 MHz to 10 GHz, CW

Accuracy: ±0.02% (Internal); External Reference capable

Amplitude Range: -40 dBm to +16 dBm

Resolution: 0.02 dB

SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset

Example: Anritsu MG3692B or equivalent

LF Counter (LFCTR):

Ch. A Frequency Range: 10 MHz

Accuracy: Based on external frequency standard

Example: Fluke PM 6690/6xx or equivalent

Frequency Standard:

Frequency Range: Suitable for use with configured LF counter

Accuracy: ±0.25 ppm or better (standard)

±0.025 ppm or better (Option FSP-B4)

Additional Equipment Required:

See "Sub FSP Add'l Equipment Minimum Use Requirements" for specific

equipment, adapters, and cables required.

The average run times during testing were:

1 hr., 45 min.