R&S FSU26: (1 yr) VER /LPNX,4419,DMM,LFCTR,SWPG,UWSG

Date: 2011-03-02

Revision: 1.2

Automated Reference: Agilent E4418B

Automated Reference: Agilent E4419B

Automated Reference: Digital Multimeter

Automated Reference: Fluke 9640A

Automated Reference: Fluke 9640A-50

Automated Reference: Fluke 9640A-LPNX

Automated Reference: HP 8485A

Automated Reference: LF Counter

Automated Reference: Low-Frequency Counter

Automated Reference: Sweep Generator

Automated Reference: µW Signal Generator

Manual Reference: Agilent 11667B

Model(s) Covered:

FSU26

Source(s):

Rohde&Schwarz Service Manual Instrument, Spectrum Analyzer, R&S®

FSU3, R&S® FSU8, R&S® FSU26, R&S® FSU31, R&S® FSU32, R&S® FSU43,

R&S® FSU46, R&S® FSU50, R&S® FSU67, Part No. 1166.1725.82-07-

Compatibility:

MET/CAL 7.3.26 or later

Flexible Standard(s) Required:

Synthesized Sweep Generator (SWPG):

Frequency Range: 10 MHz to 26.5 GHz, CW

Accuracy: ±0.02% (Internal); External Reference capable

Amplitude Range: -40 dBm to +16 dBm

Resolution: 0.05 dB

SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset

Example: HP 83630B or equivalent

µW Signal Generator (UWSG):

Frequency Range: 10 MHz to 26.5 GHz, CW

Accuracy: ±0.02% (Internal); External Reference capable

Amplitude Range: -40 dBm to +16 dBm

Accuracy: ±2 dB

SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset

Example: Anritsu MG3694B or equivalent

LF Counter (LFCTR):

Ch. A Frequency Range: 10 MHz

Accuracy: Based on external frequency standard

Example: Fluke PM 6690/6xx or equivalent

Digital Multimeter (DMM):

DC Current Function : 0 to ±10 mA

Accuracy: ±(4% + 0.25 mA) or better

Example: Fluke 8845A or equivalent

Frequency Standard:

Frequency Range: Suitable for use with configured LF counter

Accuracy: ±0.025 ppm or better (standard)

±0.0075 ppm or better (Option FSU-B4)

Additional Equipment Required:

See "Sub FSU Add'l Equipment Minimum Use Requirements" for specific

equipment, adapters, and cables required.

The average run times during testing were:

1 hr., 25 min.