R&S FSV7: (1 yr) CAL VER /LPNX,4419,LO,SWPG,VNA, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
R&S FSV7: (1 yr) CAL VER /LPNX,4419,LO,SWPG,VNA
Date: 2011-08-08
Revision: 1.1
Automated Reference: Agilent E4418B
Automated Reference: Agilent E4419B
Automated Reference: Fluke 9640A
Automated Reference: Fluke 9640A-50
Automated Reference: Fluke 9640A-LPNX
Automated Reference: HP 8481A
Automated Reference: Local Oscillator
Automated Reference: Network Analyzer
Automated Reference: Sweep Generator
Manual Reference: Agilent 11667A
Model(s) Covered:
FSV7
Source(s):
R&S® Service Manual Instrument, Spectrum Analyzer, R&S® FSV3,
R&S® FSV7, R&S® FSV13, R&S® FSV30, R&S® FSV31, R&S® FSV40, Part No.
1307.9331.82-08
R&S® FSV Spectrum Analyzer Data Sheet, Version 06.00, February 2010
Compatibility:
MET/CAL 7.3 SP1 (7.3.36) or later
Flexible Standard(s) Required:
Synthesized Sweep Generator (SWPG):
Frequency Range: 10 MHz to 10 GHz, CW
Accuracy: ±0.02% (Internal); External Reference capable
Amplitude Range: -40 dBm to +16 dBm
Resolution: 0.02 dB
SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset
Example: HP 83630B or equivalent
Local Oscillator (LO):
Frequency Range: 10 MHz to 10 GHz, CW
Accuracy: ±0.02% (Internal); External Reference capable
Amplitude Range: -40 dBm to +16 dBm
Resolution: 0.02 dB
SSB Phase Noise: <-80 dBc/Hz @ 50 kHz offset
Example: Anritsu MG3692B or equivalent
Frequency Standard:
Frequency Range: Suitable for use with configured LF counter
Accuracy: ±0.25 ppm or better (standard)
±0.025 ppm or better (Option FSV-B4)
Additional Equipment Required:
See "Sub FSV Add'l Equipment Minimum Use Requirements" for specific
equipment, adapters, and cables required.
The average run times during testing were:
1 hr., 45 min.