Package: tek_afg3102_1yr_ver_nrvs_rsa3000_tds5104b_lfctr_dmm.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Tek AFG3102:(1 yr) VER /NRVS,RSA3000,TDS5104B,LFCTR,DMM

Date: 2010-06-30 09:13:40

Revision: 1.1

Automated Reference: Digital Multimeter

Automated Reference: Low Frequency Counter

Models covered:

AFG 3102

Source(s):

AFG3000 Series Arbitrary/Function Generators Service Manual

071-1640-05

Compatibility:

MET/CAL 7.20 SP3 (7.20Z) or later

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1 based on Rohde & Schwarz NRVS and NRV-Z5 diode sensor

specifications.

AC Flatness:

1 MHZ Relative to 100 kHz TUR 2.5000

Flexible Standards Required:

LF Counter (LFCTR):

Ch. A Frequency Range: 10 MHz

Accuracy: 0.01 ppm or better

Example: Agilent 53132A or equivalent

Digital Multimeter (DMM)

AC volts: 0.1% accuracy or better

DC volts: 50 ppm, resolution 100¦ÌV accuracyor better

Resistance: ¡À 0.05 Ohm accuracy or better

Example: Agilent 3458A or equivalent

User configurated standards Required:

Power Meter/Sensor:

100 kHz to 250 MHz, 1 ¦ÌW to 100 mW(-30 dBm to +20 dBm)

Accuracy: 0.02 dB, resolution: 0.01 dB

Standard Used: Rohde & Schwarz NRVS power meter

Rohde & Schwarz NRV-Z5 power sensor

Power Sensor:

100 kHz to 250 MHz, 1 ¦ÌW to 100 mW (¨C30 dBm to +20 dBm)

Standard Used: Rohde & Schwarz NRV-Z5

Oscillioscope:

1 GHz Bandwidth, 50 Ohm input termination

Standard Used: Tektronix TDS5104B

Spectrum Analyzer:

20 kHz to 1.25 GHz

Standard Used: Tektronix RSA3000 series

Additional Accessory Required:

Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36

Adapter, BNC (f) to Type "N" (m), 50 Ohm

Adapter, BNC (f) to Dual-Banana plug

50 Ohm terminator

10X (50 Ohm) attenuator