Tektronix DMM4020: (1 year) CAL VER RS-232 /5520,8508, Revision: 1.2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Tektronix DMM4020: (1 year) CAL VER RS-232 /5520,8508
Date: 2010-04-01 13:04:26
Revision: 1.2
Automated Reference: Fluke 5520A
Automated Reference: Fluke 8508A
Manual Reference: Fluke 8508A
Model(s) Covered:
DMM4020
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1. In order to improve this, we use 8508A to characterize
the 5520A in advance for test points whose TUR which are lower than 3.5:1.
Here is list of TUR before (TUR1) and after characterization (TUR2).
DC CURRENT
- Range 200 µA: ±190 µA ; TUR1 1.38; TUR2 4.21
- Range 2000 µA: ±1900 µA ; TUR1 2.00; TUR2 8.98
- Range 200 mA: ±190 mA ; TUR1 3.40; TUR2 8.24
- Range 2 A: ±1.9 A ; TUR1 2.52; TUR2 5.48
- Range 10 A: ±10 A ; TUR1 3.82
AC CURRENT
- Range 200 mA: 190 mA, 2 kHz; TUR1 2.88; TUR2 8.77
- Range 2 A: 1.9 A, 2 kHz; TUR1 0.56; TUR2 3.69
Source:
8808A_Specs_F_Review.pdf
Testpoint_20071017.xls
Compatibility:
MET/CAL 7.20 SP1 or later
Additional Equipment Required:
4-Wire short, Fluke Part No. 2653346
The average run time during testing was 14 minutes in TAR mode.
The average run time during testing was 18 minutes in Meas. Unc. mode.