Package: tek_dmm4020_1yr_ver_rs232_5700_5725_5520.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Tek DMM4020: (1 year) CAL VER RS-232 /5700,5725,5520

Date: 2010-04-01 13:08:36

Revision: 1.2

Automated Reference: Fluke 5520A

Automated Reference: Fluke 5700A

Automated Reference: Fluke 5725A

Model(s) Covered:

DMM4020

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1 based on 5700A 1 year specifications.

2-WIRE RESISTANCE

- Range 200 Ohm: 0 Ohm ; TUR 3.9024

DC CURRENT

- Range 200 µA: ±190 µA ; TUR 3.1308

- Range 2000 µA: ±1900 µA ; TUR 3.8710

AC CURRENT

- Range 20 mA: 19 mA, 2 kHz; TUR 3.1421

- Range 200 mA: 190 mA, 2 kHz; TUR 3.7705

Source:

8808A_Specs_F_Review.pdf

Testpoint_20071017.xls

Compatibility:

MET/CAL 7.20 SP1 or later

Additional Equipment Required:

4-Wire short, Fluke Part No. 2653346

The average run time during testing was 17 minutes in TAR mode.

The average run time during testing was 20 minutes in Meas. Unc. mode.