Tek DMM4040: (1 yr) VER IEEE /5720,5725,5520,8508, Revision: 1.2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Tek DMM4040: (1 yr) VER IEEE /5720,5725,5520,8508
Date: 2010-09-02 10:55:27
Revision: 1.2
Automated Reference: Fluke 5520A
Automated Reference: Fluke 5720A
Automated Reference: Fluke 5725A
Automated Reference: Fluke 8508A
Model Covered:
DMM4040
This procedure contains the following tests whose Test Uncertainty Ratios
are less than 4:1. With the exception of 1 GOhm, a guardbanding strategy
(chosen by the operator at run time) is used to compensate for the low
TURs. In the case of 1 GOhm, the 5520A 1 GOhm output is characterized
immediately prior to use, using an 8508A.
The following are the TURs less than 4:1:
AC VOLTAGE
- Range 100 mV: 100 mV, 10 Hz; TUR 2.22
- Range 1 V: 1 V, 10 Hz; TUR 2.57
- Range 10 V: 10 V, 10 Hz; TUR 2.57
- Range 1000 V: 320 V, 50 kHz; TUR 3.74
- Range 1000 V: 320 V, 100 kHz; TUR 3.23
Source:
Fluke 8845A/8846A Digital Multimeter Calibration Manual
January 2007, Rev. 1, 11/07
Fluke 8845A/8846A Digital Multimeter Calibration Manual Supplement
Revision/Date: 1, 11/07, Supplement Issue: 5, Issue Date: 4/10
Tektronix DMM4040 and DMM4050 Digital Multimeter Users Manual
077-0361-00
Compatibility:
MET/CAL 7.20 SP3 (7.20Z) or later
Additional Equipment Required:
See "Sub Fluke 8845A/46A VER Required Equipment"
The average run time during testing was 38 minutes.