Tek DMM4040: (1 yr) VER IEEE /5720,5725,5520,8508

Date: 2010-09-02 10:55:27

Revision: 1.2

Automated Reference: Fluke 5520A

Automated Reference: Fluke 5720A

Automated Reference: Fluke 5725A

Automated Reference: Fluke 8508A

Model Covered:

DMM4040

This procedure contains the following tests whose Test Uncertainty Ratios

are less than 4:1. With the exception of 1 GOhm, a guardbanding strategy

(chosen by the operator at run time) is used to compensate for the low

TURs. In the case of 1 GOhm, the 5520A 1 GOhm output is characterized

immediately prior to use, using an 8508A.

The following are the TURs less than 4:1:

AC VOLTAGE

- Range 100 mV: 100 mV, 10 Hz; TUR 2.22

- Range 1 V: 1 V, 10 Hz; TUR 2.57

- Range 10 V: 10 V, 10 Hz; TUR 2.57

- Range 1000 V: 320 V, 50 kHz; TUR 3.74

- Range 1000 V: 320 V, 100 kHz; TUR 3.23

Source:

Fluke 8845A/8846A Digital Multimeter Calibration Manual

January 2007, Rev. 1, 11/07

Fluke 8845A/8846A Digital Multimeter Calibration Manual Supplement

Revision/Date: 1, 11/07, Supplement Issue: 5, Issue Date: 4/10

Tektronix DMM4040 and DMM4050 Digital Multimeter Users Manual

077-0361-00

Compatibility:

MET/CAL 7.20 SP3 (7.20Z) or later

Additional Equipment Required:

See "Sub Fluke 8845A/46A VER Required Equipment"

The average run time during testing was 38 minutes.