Package: tek_dmm4040_1yr_ver_ieee_5720_5725_fgen_9331s.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Tek DMM4040: (1 yr) VER IEEE /5720,5725,FGEN,9331S/1G, Revision: 1.2
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Tek DMM4040: (1 yr) VER IEEE /5720,5725,FGEN,9331S/1G
Date: 2011-02-16 14:57:51
Revision: 1.2
Automated Reference: Fluke 5720A
Automated Reference: Fluke 5725A
Automated Reference: Function Generator
Manual Reference: MIL 9331S/1G
Model Covered:
DMM4040
This procedure contains the following tests whose Test Uncertainty Ratios
are less than 4:1. A guardbanding strategy (chosen by the operator at run
time) is used to compensate for the low TURs.
AC VOLTAGE
- Range 100 mV: 100 mV, 10 Hz; TUR 2.38
- Range 1 V: 1 V, 10 Hz; TUR 2.81
- Range 10 V: 10 V, 10 Hz; TUR 2.81
- Range 1000 V: 320 V, 50 kHz; TUR 3.74
- Range 1000 V: 320 V, 100 kHz; TUR 3.23
Source:
Fluke 8845A/8846A Digital Multimeter Calibration Manual
January 2007, Rev. 1, 11/07
Fluke 8845A/8846A Digital Multimeter Calibration Manual Supplement
Revision/Date: 1, 11/07, Supplement Issue: 5, Issue Date: 4/10
Tektronix DMM4040 and DMM4050 Digital Multimeter Users Manual
077-0361-00
Compatibility:
MET/CAL 7.20 SP3 (7.20Z) or later
Additional Equipment Required:
See "Sub Fluke 8845A/46A VER Required Equipment"
The average run time during testing was 38 minutes.