Package: tek_dmm4040_1yr_ver_ieee_5720_5725_fgen_9331s.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Tek DMM4040: (1 yr) VER IEEE /5720,5725,FGEN,9331S/1G

Date: 2011-02-16 14:57:51

Revision: 1.2

Automated Reference: Fluke 5720A

Automated Reference: Fluke 5725A

Automated Reference: Function Generator

Manual Reference: MIL 9331S/1G

Model Covered:

DMM4040

This procedure contains the following tests whose Test Uncertainty Ratios

are less than 4:1. A guardbanding strategy (chosen by the operator at run

time) is used to compensate for the low TURs.

AC VOLTAGE

- Range 100 mV: 100 mV, 10 Hz; TUR 2.38

- Range 1 V: 1 V, 10 Hz; TUR 2.81

- Range 10 V: 10 V, 10 Hz; TUR 2.81

- Range 1000 V: 320 V, 50 kHz; TUR 3.74

- Range 1000 V: 320 V, 100 kHz; TUR 3.23

Source:

Fluke 8845A/8846A Digital Multimeter Calibration Manual

January 2007, Rev. 1, 11/07

Fluke 8845A/8846A Digital Multimeter Calibration Manual Supplement

Revision/Date: 1, 11/07, Supplement Issue: 5, Issue Date: 4/10

Tektronix DMM4040 and DMM4050 Digital Multimeter Users Manual

077-0361-00

Compatibility:

MET/CAL 7.20 SP3 (7.20Z) or later

Additional Equipment Required:

See "Sub Fluke 8845A/46A VER Required Equipment"

The average run time during testing was 38 minutes.