Tek DMM4050:(1 yr)VER IEEE/5720,5725,FGEN,1409,9331S/1G

Date: 2011-02-16 14:57:56

Revision: 1.2

Automated Reference: Fluke 5720A

Automated Reference: Fluke 5725A

Automated Reference: Function Generator

Manual Reference: GenRad 1409-F

Manual Reference: GenRad 1409-L

Manual Reference: GenRad 1409-T

Manual Reference: GenRad 1409-Y

Manual Reference: IET 1409-1000uF

Manual Reference: IET 1409-100uF

Manual Reference: IET 1409-10uF

Manual Reference: MIL 9331S/1G

Model Covered:

DMM4050

This procedure contains the following tests whose Test Uncertainty Ratios

are less than 4:1. A guardbanding strategy (chosen by the operator at run

time) is used to compensate for the low TURs.

AC VOLTAGE

- Range 100 mV: 100 mV, 10 Hz; TUR 2.38

- Range 1 V: 1 V, 10 Hz; TUR 2.81

- Range 10 V: 10 V, 10 Hz; TUR 2.81

- Range 1000 V: 320 V, 50 kHz; TUR 3.74

- Range 1000 V: 320 V, 100 kHz; TUR 3.23

CAPACITANCE

- Range 1 nF: 1 nF ; TUR 3.00

- Range 100 µF: 100 µF ; TUR 2.73

- Range 1 mF: 1.0 mF ; TUR 2.73

- Range 10 mF: 10 mF ; TUR 2.73

- Range 1 mF: 100 mF ; TUR 3.50

Source:

Fluke 8845A/8846A Digital Multimeter Calibration Manual

January 2007, Rev. 1, 11/07

Fluke 8845A/8846A Digital Multimeter Calibration Manual Supplement

Revision/Date: 1, 11/07, Supplement Issue: 5, Issue Date: 4/10

Tektronix DMM4040 and DMM4050 Digital Multimeter Users Manual

077-0361-00

Compatibility:

MET/CAL 7.20 SP3 (7.20Z) or later

Additional Equipment Required:

See "Sub Fluke 8845A/46A VER Required Equipment"

The average run time during testing was 38 minutes.