Package: tek_dpo4000_1yr_ver_visa_5520_s6_s11.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Tektronix DPO 4032: (1 year) CAL VER VISA /5520+SC600, Revision: 1.5
Tektronix DPO 4034: (1 year) CAL VER VISA /5520+SC600, Revision: 1.5
Tektronix DPO 4054: (1 year) CAL VER VISA /5520+SC600, Revision: 1.5
Tektronix DPO 4104: (1 year) CAL VER VISA /5520+SC1100, Revision: 1.5
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Tektronix DPO 4032: (1 year) CAL VER VISA /5520+SC600
Date: 2011-08-23 11:24:35
Revision: 1.5
Automated Reference: Fluke 5520A
Tektronix DPO 4034: (1 year) CAL VER VISA /5520+SC600
Tektronix DPO 4054: (1 year) CAL VER VISA /5520+SC600
Date: 2011-08-23 11:24:39
Revision: 1.5
Automated Reference: Fluke 5520A
Models covered:
DPO 4034, DPO 4054
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1.
OFFSET ACCURACY
50 Ohm, 1 mV/div; 0.9 V Input/Offset; TUR 2.82
50 Ohm, 1 mV/div; -0.9 V Input/Offset; TUR 2.82
50 Ohm, 2 mV/div; 0.5 V Input/Offset; TUR 3.00
50 Ohm, 2 mV/div; -0.5 V Input/Offset; TUR 3.00
Source(s):
MSO4000 and DPO4000 Series Digital Phosphor Oscilloscopes
Specification and Performance Verification Technical Reference
PN 077-0247-00
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36
50 Ohm feed through teminator
Tektronix DPO 4104: (1 year) CAL VER VISA /5520+SC1100
Date: 2011-08-23 11:24:42
Revision: 1.5
Automated Reference: Fluke 5520A
Models covered:
DPO 4104
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1.
OFFSET ACCURACY
50 Ohm, 1 mV/div; 0.9 V Input/Offset; TUR 2.82
50 Ohm, 1 mV/div; -0.9 V Input/Offset; TUR 2.82
50 Ohm, 2 mV/div; 0.5 V Input/Offset; TUR 3.00
50 Ohm, 2 mV/div; -0.5 V Input/Offset; TUR 3.00
Source(s):
MSO4000 and DPO4000 Series Digital Phosphor Oscilloscopes
Specification and Performance Verification Technical Reference
PN 077-0247-00
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36
50 Ohm feed through teminator
Models covered:
DPO 4032
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1.
OFFSET ACCURACY
50 Ohm, 1 mV/div; 0.9 V Input/Offset; TUR 2.82
50 Ohm, 1 mV/div; -0.9 V Input/Offset; TUR 2.82
50 Ohm, 2 mV/div; 0.5 V Input/Offset; TUR 3.00
50 Ohm, 2 mV/div; -0.5 V Input/Offset; TUR 3.00
Source(s):
MSO4000 and DPO4000 Series Digital Phosphor Oscilloscopes
Specification and Performance Verification Technical Reference
PN 077-0247-00
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Additional Equipment Required:
Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36
50 Ohm feed through teminator