Package: tek_dpo7104_1yr_ver_ieee_5800_ghz_5ch_tdp_lfctr.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Tek DPO 7104:(1 yr) CAL VER IEEE/5800+5Ch+GHz+TDP,LFCTR, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Tek DPO 7104:(1 yr) CAL VER IEEE/5800+5Ch+GHz+TDP,LFCTR
Date: 2011-02-04 11:12:02
Revision: 1.1
Automated Reference: Fluke 5800A
Automated Reference: Low Frequency Counter
Automated Reference: Low-Frequency Counter
Manual Reference: Fluke 5800A/TDP
Models covered:
DPO 7104
This procedure contains the following tests whose Test Uncertainty Ratio
are less than 4:1.
DCV ACCURACY
50 Ohm, 1 mV/div, Positive; TUR 2.1033
50 Ohm, 1 mV/div, Negative; TUR 2.1033
50 Ohm, 2 mV/div, Positive; TUR 2.2707
50 Ohm, 2 mV/div, Negative; TUR 2.2707
50 Ohm, 5 mV/div, Positive; TUR 2.6906
50 Ohm, 5 mV/div, Negative; TUR 2.6906
50 Ohm, 10 mV/div, Positive; TUR 3.8926
50 Ohm, 10 mV/div, Negative; TUR 3.8926
50 Ohm, 50 mV/div, Positive; TUR 2.6201
50 Ohm, 50 mV/div, Negative; TUR 2.6201
Source(s):
DPO7000, DSA/DPO70000, and DSA/DPO7000B Series Digital Phosphor
Oscilloscopes Specification and Performance Verification
Technical Reference PN 077-0063-01
Compatibility:
MET/CAL 7.20 SP1 (7.20x) or later
Flexible Standard(s) Required:
LF Counter (LFCTR):
Ch. A Frequency Range: 10 MHz
Accuracy: ±3.5 ppm/year or better
Example: Fluke PM6681 with PM9678B option or equivalent
Additional Equipment Required:
Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36
One BNC to Minigrabber adapter
One 50 Ohm, 60 inch presision coaxial cable, male-to-male
SMA connectors.
One BNC elbow connector
One SMA "T", male to two SMA female connectors
One SMA female to BNC male connector
One SMA termination connector, short circuit.
One SMA male-to-female BNC adapter
One attenuator, 10X, impedance 50 Ohm, BNC <= 2 GHz; SMA <= 18 GHz
One attenuator, 5X, impedance 50 Ohm, BNC <= 2 GHz; SMA <= 18 GHz
One 50 Ohm terminator
One power splitter, frequency range: DC to 18 GHz, tracking < 2.0%
One attenuator, 2X, impedance 50 Ohm.