Package: tek_dpo7104_1yr_ver_ieee_9500_1100_4hd_lfctr.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Tek DPO 7104: (1 yr) CAL VER IEEE /9500/1100+4Hd,LFCTR

Date: 2011-04-29 11:41:56

Revision: 1.4

Automated Reference: Datron 9500

Automated Reference: Datron 9510

Automated Reference: Datron 9520

Automated Reference: Low Frequency Counter

Models covered:

DPO 7104

Source(s):

DPO7000, DSA/DPO70000, and DSA/DPO7000B Series Digital Phosphor

Oscilloscopes Specification and Performance Verification

Technical Reference PN 077-0063-01

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

Flexible Standard(s) Required:

LF Counter (LFCTR):

Ch. A Frequency Range: 10 MHz

Accuracy: ±3.5 ppm/year or better

Example: Fluke PM6681 with PM9678B option or equivalent

Additional Equipment Required:

Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36

One 2X attenuator

One BNC to Minigrabber adapter

Adapter SMA(f)-BNC(m), one to four sets(if 9560 configured)

One 50 Ohm, 60 inch presision coaxial cable, male-to-male

SMA connectors.

One BNC elbow connector

One SMA "T", male to two SMA female connectors

One SMA female to BNC male connector

One SMA termination connector, short circuit.

One SMA male-to-female BNC adapter