Tektronix MSO 3014: (1 year) VER VISA /5800,DMM

Date: 2010-12-03 12:28:19

Revision: 1.4

Automated Reference: Digital Multimeter

Automated Reference: Fluke 5800A

Models covered:

MSO 3014

Source(s):

MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes

Specification and Performance Verification Manual, PN 077-0300-00

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

DCV GAIN:

1 MOhm, 1 mV/div TUR 1.43

1 MOhm, 2 mV/div TUR 2.78

Additional Equipment Required:

Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36

50 Ohm terminator

Tek P6316 Logic Probe 16 Channels

BNC-to-0.1 inch pin adapter