Tektronix TDS 3014C: (1 yr) CAL VER VISA /9100+Opt600

Date: 2010-06-04 09:18:40

Revision: 1.3

Automated Reference: Datron 9100

Models covered:

TDS 3014C

Source(s):

TDS3000C Series Digital Phosphor Oscilloscopes,

Specifications and Performance Verification

Technical Reference, 071-2508-01

Compatibility:

MET/CAL 7.20 SP1 (7.20x) or later

This procedure contains the following tests whose Test Uncertainty Ratio

are less than 4:1.

DC VOLTAGE ACCURACY

1 mV/div, 100 mV input TUR 3.1250

Additional Equipment Required:

Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-36