Package: tek_ths3014_1yr_ver_rs232_5500_sc600.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Tektronix THS3014: (1 year) CAL VER RS-232 /5500+SC600

Date: 2012-11-07 13:17:57

Revision: 1.1

Automated Reference: Fluke 5500A

Models covered:

THS3014

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5500A (1 year) specifications.

INPUT A AC VOLTAGE ACCURACY (HF) & BANDWIDTH TEST

900 mV @ 1 MHz; TUR 2.0566

INPUT B AC VOLTAGE ACCURACY (HF) & BANDWIDTH TEST

900 mV @ 1 MHz; TUR 2.0566

INPUT C AC VOLTAGE ACCURACY (HF) & BANDWIDTH TEST

900 mV @ 1 MHz; TUR 2.0566

INPUT D AC VOLTAGE ACCURACY (HF) & BANDWIDTH TEST

900 mV @ 1 MHz; TUR 2.0566

Source:

Tektronix THS3000 Series Oscilloscopes

Performance Verification and Adjustments Technical Reference

PN 077-0606-00

Compatibility:

MET/CAL or 5500/CAL 7.3 SP1 (7.3.36) or later

Required Equipment:

Adapter, dual banana plug to BNC(f), Pomona 1269

Adapter, dual BNC(f) to BNC(m), Fluke PM 9093, 3 ea

Termination, feedthrough, BNC, 50 Ohm, Fluke PM 9585

Cable, 50 Ohm, BNC(m)-(m), Pomona 2249-C-6, 1 ea.

Cable, USB Type A (Host) to USB MINI Type B (Device)

Accessories supplied with UUT

- Power Adapter, BC190

- THP0301 Voltage Probe (200 MHz, 10:1), any color

The average run time during testing was:

60 minutes (using procedure default values for NTHROW and NMEAS)