Package: tek_tx3_1yr_ver_5520.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Tektronix TX3: (1 year) CAL VER RS-232 /5520

Date: 2011-06-24 10:55:37

Revision: 2

Automated Reference: Fluke 5520A

Model(s) Covered:

TX3

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 5520A (1 year) specifications.

CAPACITANCE TEST

- 4 nF; TUR 2.2500

- 40 nF; TUR 2.1500

- 400 nF; TUR 2.4432

- 4 µF; TUR 2.1500

- 40 mF; TUR 2.2778

AC CURRENT TEST

- 460 mA, 3 kHz; TUR 1.2766

- 460 mA, 5 kHz; TUR 2.5000

Source:

Tektronix TX1/TX3 Service Manual, Part No. 070-9893-01 (No date or

revision stated)

Compatibility:

MET/CAL or 5500/CAL 7.01 SP2 or later

Additional Equipment Required:

DB9 IR Cable Kit, Fluke Part No. 1590638

Capacitance Calibration Cable, Fluke Part No. 1612912