Wavetek 2015: (1 year) CAL VER /5520
Date: 2005-02-08
Revision: 1.1
Automated Reference: Fluke 5520A
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 5520A (1 year) specifications.
Capacitance Tests:
4 nF Range, 3.500nF; TUR 4.0000
40 nF Range, 35.00nF; TUR 2.0800
400 nF Range, 350.0nF; TUR 2.0800
4 µF Range, 3.500uF; TUR 2.0800
40 µF Range, 10.00uF; TUR TUR 4.0000
Model(s) Covered:
Wavetek 2015
Source:
Wavetek 2015 Operator's Manual, part number 1566149
Compatibility:
MET/CAL or 5500/CAL 7.01 SP2 or later
Additional Equipment Required:
None
The average run time during testing was 15 minutes (keyboard entry).