Package: wavetek_hd140b_1yr_ver_9100.mcz
Become a MET/SUPPORT Gold Member to download this package!
Procedures in Package
Click on the procedure names to find out details about the procedure files containing them.
Beckman Industrial HD140B: (1 year) CAL VER /9100, Revision: 1.1
Wavetek HD140B: (1 year) CAL VER /9100, Revision: 1.1
Details about Procedure Files
Below is a list of procedure files (.pro files) in the package.
Subprocedures and procedures files that only contain subprocedures have been excluded.
The references of the subprocedure files have been included.
Beckman Industrial HD140B: (1 year) CAL VER /9100
Wavetek HD140B: (1 year) CAL VER /9100
Date: 2007-05-18 09:13:40
Revision: 1.1
Automated Reference: Datron 9100
This procedure contains tests whose Test Uncertainty Ratios are less
than 4:1, based on 9100 (1 year) specifications.
AC CURRENT
200 µA Range, 190.0uA @45Hz; TUR 3.5335
RESISTANCE TESTS:
2 MOhm Range, 1.900 MZ; TUR 2.0000
20 MOhm Range, 19.00 MZ; TUR 2.5246
Model(s) Covered:
Wavetek / Beckman Industrial HD140B
Source:
Wavetek HD140B Operator's Manual
Compatibility:
MET/CAL 7.01 SP2 or later
Additional Equipment Required:
None
The average run time during testing was 10 minutes (keyboard entry).