Package: wavetek_hd140b_1yr_ver_9100.mcz
Become a MET/SUPPORT Gold Member to download this package!

Details about Procedure Files

Below is a list of procedure files (.pro files) in the package.

Subprocedures and procedures files that only contain subprocedures have been excluded.

The references of the subprocedure files have been included.

Beckman Industrial HD140B: (1 year) CAL VER /9100

Wavetek HD140B: (1 year) CAL VER /9100

Date: 2007-05-18 09:13:40

Revision: 1.1

Automated Reference: Datron 9100

This procedure contains tests whose Test Uncertainty Ratios are less

than 4:1, based on 9100 (1 year) specifications.

AC CURRENT

200 µA Range, 190.0uA @45Hz; TUR 3.5335

RESISTANCE TESTS:

2 MOhm Range, 1.900 MZ; TUR 2.0000

20 MOhm Range, 19.00 MZ; TUR 2.5246

Model(s) Covered:

Wavetek / Beckman Industrial HD140B

Source:

Wavetek HD140B Operator's Manual

Compatibility:

MET/CAL 7.01 SP2 or later

Additional Equipment Required:

None

The average run time during testing was 10 minutes (keyboard entry).