Page 19 - Pressure Calibration Brochure
P. 19
50%
1
000%
%
2000 PSI 148 BAR
TECH
TIPS
• Use a reference gauge with
better accuracy to meet test
uncertainty ratios over a
wider range of pressures.
• Forgo wrenches or PTFE
tape by using adapters to
it multiple sizes and types
of devices with leak tight
seals to 20,000 psi.
• Safety irst! Always use
ittings, tubing, and seals
with pressure ratings above
full scale of the instrument.
To perform the test:
• If possible use oil for better
lubrication.
STEP
The pressure gauge should be mounted in the same orientation
(vertical or horizontal) as in the process. An angle adapter such as the 1
• Use gas to improve
P5543 may be used.
cleanliness or a liquid-to-
liquid separator available
The reference pressure gauge (2700G) should be mounted such that the STEP
from Fluke.
display is easily seen.
2
• Hydraulic systems are
preferable to gas systems
for pressures above
For hydraulic comparators prime the luid with the priming pump, STEP
2000 psi due to safety
to remove any bubbles.
3
and ease of use.
Measurement points should be distributed uniformly over the calibration STEP
range. Conveniently source pressure with a manual pump up to 300 psi, 4
after that use an external pressure supply.
For gas comparators use the ine needle valve or ine adjustment screw STEP
press to precisely meter the pressure.
5
With hydraulic models use the screw press to source and ine adjust STEP
6
the pressure.
The source pressure can be adjusted until the device under test is
STEP
reading a nominal pressure or until the reference gauge reads the 7
nominal pressure.