Page 19 - Pressure Calibration Brochure
P. 19






50%
1
000%
%



2000 PSI 148 BAR
TECH 


TIPS



• Use a reference gauge with 

better accuracy to meet test 
uncertainty ratios over a 

wider range of pressures.

• Forgo wrenches or PTFE 
tape by using adapters to 

it multiple sizes and types 
of devices with leak tight 

seals to 20,000 psi.

• Safety irst! Always use 
ittings, tubing, and seals 

with pressure ratings above 
full scale of the instrument.
To perform the test:
• If possible use oil for better 

lubrication.
STEP
The pressure gauge should be mounted in the same orientation 
(vertical or horizontal) as in the process. An angle adapter such as the 1
• Use gas to improve 

P5543 may be used.
cleanliness or a liquid-to- 
liquid separator available 

The reference pressure gauge (2700G) should be mounted such that the STEP
from Fluke.

display is easily seen.
2
• Hydraulic systems are 
preferable to gas systems 

for pressures above
For hydraulic comparators prime the luid with the priming pump, STEP
2000 psi due to safety 
to remove any bubbles.
3
and ease of use.


Measurement points should be distributed uniformly over the calibration STEP
range. Conveniently source pressure with a manual pump up to 300 psi, 4

after that use an external pressure supply.



For gas comparators use the ine needle valve or ine adjustment screw STEP
press to precisely meter the pressure.
5



With hydraulic models use the screw press to source and ine adjust STEP
6
the pressure.


The source pressure can be adjusted until the device under test is 
STEP
reading a nominal pressure or until the reference gauge reads the 7
nominal pressure.