Calibrating a HART


Smart Pressure Transmitter




Pressure transmitter manufacturers have improved the accuracy 
and technology designed into these smart pressure measurement 

devices. Many conventional calibration tools have become inad- 
equate or simply unable to test and calibrate these high accuracy 

pressure transmitters. Better test solutions are required.

Verifying and documenting the performance and adjusting
a HART smart pressure transmitter can require a bucket full of 

tools. Performing this task with a HART enabled calibrator like 
the Fluke 754 simpliies the task and reduces what you need 

to carry.

Before going to the ield: install the pressure module adapter 
to the hand pump with thread seal. Once the adapter is properly 

installed on the pump, changing modules to different pressure 
ranges is a snap, no tools required.

To get the accuracy needed: to test these new high accuracy 

transmitters match the pressure measurement standard range 
closely to the device tested. For example, use a 100 psi

pressure module to calibrate and test a transmitter ranged at 
100 psi. Industry standards suggest the measurement standard 

should be 4-10 times more accurate than the device being tested 
so best-in-class accuracy is required.

The Fluke 754 utilizes the 750P series pressure modules and 
has built-in HART functionality to enable smart trims on trans- 

mitters. It can also document transmitter performance before and 
after adjustment and calculate pass/fail errors.











Suggested test tools





















Fluke 754 Docu- Fluke 700G Fluke 750P Series Fluke 700PTP-1 
menting Process Precision Pressure Pressure Modules
Pneumatic Test Pump

Calibrator-HART
Gauge Calibrator
See pg 23
See pg 31

See pg 23
See pg 26