Page 76 - Fluke Test Tools Catalog
P. 76
Fluke current clamps
For use with multimeters, portable oscilloscopes and power quality analyzers
AC/DC current clamp specifications
AC/DC models
80i-110s Especially useful for low i30s current measurements
i310s, i30s and i30 Hall Effect AC/DC Current Clamps
i410 i310s
i1010
down to 30 milliamps.
i30
80i-110s
i410
i1010
i30
i30s
i310s
DC range
0.1 A to 100 A
1 A to 400 A
1 A to 1000 A
30 mA to 30 A
30 mA to 30 A
0.1 A to 450 A
AC range
0.1 A to 70 A
1 A to 400 A
1 A to 600 A
30 mA to 20 A rms
30 mA to 20 A rms
0.1 A to 300 A
Accuracy
0.1 A to 10 A; dc to
1 kHz: ± 3 % + 50 mA; up to 100 A add 15 %; up to 20 kHz add 12 %
3.5 % + 0.5 A for dc or ac (45 to 400 Hz)
2 % + 0.5 A for dc or ac (45 to 400 Hz)
±1% ofreading,±2mA
± 1 % of reading, ± 2 mA
30 A range: 1 % + 50 mA 300 A range: 1 % + 300 mA
Bandwidth (-3 dB)
100 kHZ
3 kHz
10 kHz
dc to 20 kHz (-0.5 dB)
dc to 100 kHz (-0.5 dB)
DC to 20 kHz
Zero error adjustment
•
•
•
•
•
•
Maximum conductor diameter
11.8 mm (0.46 in)
30 mm (1.18 in)
30 mm (1.18 in)
19 mm (0.75 in)
19 mm (0.75 in)
19 mm (0.75 in)
Maximum conductor size
1 AWG
750 MCM or 2-500 MCM
750 MCM or 2-500 MCM
250 MCM
250 MCM
250 MCM
Output levels
10 mV/A, 100 mV/A
1 mV/A
1 mV/A
100 mV/A
100 mV/A
1 mV/A, 10 mV/A
Output cable
1.6 meter with BNC termination
1.2 meter with shrouded banana plugs
1.2 meter with shrouded banana plugs
1.5 meter with shrouded dual banana plugs
2 meter with BNC termination
2 meters with BNC termination and a BNC to safety shrouded banana adapter
Battery life
40 hours
60 hours
60 hours
30 hours
30 hours
30 hours
Warranty
One-year
One-year
One-year
One-year
One-year
One-year
Safety
CAT II 600 V; CAT III 300 V
CAT III 600 V
CAT III 600 V
CAT III 300 V
CAT III 300 V
CAT III 300 V
AC/DC current clamp selection guide
Meter
80i-110s
i410
i1010
i30
i30s
i310s
114/115/116/117
0.1 A to 100 A dc/ 0.1 A to 70 A ac*
1.4 A to 400 A dc/ 6 A to 400 A ac
1.4 A to 1000 A dc/ 6 A to 600 A ac
30 mA to 30 A dc/ 60 mA to 20 A ac
30 mA to 30 A dc/ 60 mA to 20 A ac
0.1 A to 450 A dc/ 0.6 A to 300 A ac
233
0.1 A to 100 A dc/ 0.1 A to 70 A ac*
1.4 A to 400 A dc/ 6 A to 400 A ac
1.4 A to 1000 A dc/ 6 A to 600 A ac
30 mA to 30 A dc/ 60 mA to 20 A ac
30 mA to 30 A dc/ 60 mA to 20 A ac
0.1 A to 450 A dc/ 0.6 A to 300 A ac
27II
0.1 A to 100 A dc/ 0.1 A to 70 A ac*
1 A to 400 A dc/ 1 A to 400 A ac
1 A to 1000 A dc/ 1 A to 600 A ac
30 mA to 30 A dc/ 30 mA to 20 A ac
30 mA to 30 A dc/ 30 mA to 20 A ac
0.1 A to 450 A dc/ 0.1 A to 300 A ac
28II
0.1 A to 100 A dc/ 0.2 A to 70 A ac*
1 A to 400 A dc/ 6 A to 400 A ac
1 A to 1000 A dc/ 6 A to 600 A ac
30 mA to 30 A dc/ 60 mA to 20 A ac
30 mA to 30 A dc/ 60 mA to 20 A ac
0.1 A to 450 A dc/ 2 A to 300 A ac
77-IV
0.1 A to 100 A dc/ 0.1 A to 70 A ac*
1 A to 400 A dc/ 5 A to 400 A ac
1 A to 1000 A dc/ 5 A to 600 A ac
30 mA to 30 A dc/ 40 mA to 20 A ac
30 mA to 30 A dc/ 40 mA to 20 A ac
0.1 A to 450 A dc/ 0.4 A to 300 A ac
175/177/179
0.11 A to 100 A dc/ 0.3 A to 70 A ac*
1.6 A to 400 A dc/ 30 A to 400 A ac
1.6 A to 1000 A dc/ 30 A to 600 A ac
30 mA to 30 A dc/ 300 mA to 20 A ac
30 mA to 30 A dc/ 300 mA to 20 A ac
0.1 A to 450 A dc/ 3 A to 300 A ac
1577/1587 Insulation Meter
0.1 A to 100 A dc/ 0.3 A to 70 A ac*
1.2 A to 400 A dc/ 30 A to 400 A ac
1.2 A to 1000 A dc/ 30 A to 600 A ac
30 mA to 30 A dc/ 300 mA to 20 A ac
30 mA to 30 A dc/ 300 mA to 20 A ac
0.1 A to 450 A dc/ 3 A to 300 A ac
83V/88V
0.1 A to 100 A dc/ 0.1 A to 70 A ac*
1 A to 400 A dc/ 1 A to 400 A ac
1 A to 1000 A dc/ 1 A to 600 A ac
30 mA to 30 A dc/ 30 mA to 20 A ac
30 mA to 30 A dc/ 30 mA to 20 A ac
0.1 A to 450 A dc/ 0.1 A to 300 A ac
87V
0.1 A to 100 A dc/ 0.2 A to 70 A ac*
1 A to 400 A dc/ 20 A to 400 A ac
1 A to 1000 A dc/ 20 A to 600 A ac
30 mA to 30 A dc/ 200 mA to 20 A ac
30 mA to 30 A dc/ 200 mA to 20 A ac
0.1 A to 450 A dc/ 2 A to 300 A ac
287/289
0.1 A to 100 A dc/ 0.1 A to 70 A ac*
1 A to 400 A dc/ 1 A to 400 A ac
1 A to 1000 A dc/ 1 A to 600 A ac
30 mA to 30 A dc/ 30 mA to 20 A ac
30 mA to 30 A dc/ 30 mA to 20 A ac
0.1 A to 450 A dc/ 0.1 A to 300 A ac
123/124 ScopeMeters
0.1 A to 100 A dc/ 0.25 A to 70 A ac
1 A to 400 A dc/ 25 A to 400 A ac**
1 A to 1000 A dc/ 25 A to 600 A ac**
30 mA to 30 A dc/ 250 mA to 20 A ac**
30 mA to 30 A dc/ 250 mA to 20 A ac
0.1 A to 450 A dc/ 2.5 A to 300 A ac
190 Series ScopeMeters
0.1 A to 100 A dc/ 0.1 A to 70 A ac
1 A to 400 A dc/ 5 A to 400 A ac**
1 A to 1000 A dc/ 5 A to 600 A ac**
30 mA to 30 A dc/ 50 mA to 20 A ac**
30 mA to 30 A dc/ 50 mA to 20 A ac
0.1 A to 450 A dc/ 0.5 A to 300 A ac
8808A
0.1 A to 100 A dc/ 0.1 A to 70 A ac*
1 A to 400 A dc/ 2 A to 400 A ac
1 A to 1000 A dc/ 2 A to 600 A ac
30 mA to 30 A dc/ 30 mA to 20 A ac
30 mA to 30 A dc/ 30 mA to 20 A ac
0.1 A to 450 A dc/ 0.2 A to 300 A ac
8845A/8846A
0.1 A to 100 A dc/ 0.1 A to 70 A ac*
1 A to 400 A dc/ 1 A to 400 A ac
1 A to 1000 A dc/ 1 A to 600 A ac
30 mA to 30 A dc/ 30 mA to 20 A ac
30 mA to 30 A dc/ 30 mA to 20 A ac
0.1 A to 450 A dc/ 0.1 A to 300 A ac
705/707
0.1 A to 100 A dc, no ac*
4 A to 400 A dc, no ac
4 A to 1000 A dc, No ac
40 mA to 30 A dc, no ac
40 mA to 30 A dc, no ac
0.4 A to 450 A dc, no ac
715
0.1 A to 100 A dc, no ac*
4 A to 400 A dc, no ac
4 A to 1000 A dc, No ac
40 mA to 30 A dc, no ac
40 mA to 30 A dc, no ac
0.1 A to 450 A dc, no ac
787/789 Process Meters
0.1 A to 100 A dc/ 0.2 A to 70 A ac*
1 A to 400 A dc/ 20 A to 400 A ac
1 A to 1000 A dc/ 20 A to 600 A ac
30 mA to 30 A dc/ 200 mA to 20 A ac
3 mA to 30 A dc/ 200 mA to 20 A ac
0.1 A to 450 A dc/ 2 A to 300 A ac
74
Accessories
For more information, go to www.fluke.com/accessories 74
* Requires PM9081 Adapter
** Requires PM9082/001 Adapter
***Only for use with the 430 Series Power Quality Analyzers